Criticare Multiparameter Patient Monitors
Scholar I
new model

Scholar I Specifications
The following Scholar™ product specifications are subject to change without notice:
ECG
Number of Leadwires 3 or 5 (selectable)
Lead Selection I, II, III, aVR, aVL, aVF, V
Gain Selection 0.5 mV, 1 mV, 2 mV, 4 mV full scale
(all ±2%)
Frequency Response Diagnostic: 0.05-100 Hz (-3db)
Monitor: 0.50-40 Hz (-3db)
Isolation Conforms to IEC 601-1
Sweep Speed 6.25, 12.5, 25, 50 mm/sec
Electrosurgery Protection Yes
Defibrillation Protection Yes
Pacer Rejection Yes
Heart Rate Display and Alarms
Range 20-300 bpm
Update Time Every beat
Accuracy ± 1 bpm or 1% (the larger)
Alarm Limit Range 80-300 bpm (high) 20-150 bpm (low)
Alarm Indication Audio, visual
Alarm Silence Yes
Non-Invasive Blood Pressure
Technique Oscillometric upon inflation
Measurement Cycle (0-300 mmHg: std. adult cuff)
<40 sec average Automatic Measurement Cycles 1, 2, 3, 4, 5, 10, 15, 30, 60 min Range of Pressure 30-300 mmHg Adult 20-150 mmHg Neonate Accuracy Greater of ±2% or ±2mmHg over full range Invasive Pressure (optional)
Isolation Breakdown Voltage 4kV RMS, 5.5 kV peak from ground-related currents
Leakage current 10µA
Pressure Range -10 to 350 mmHg
Zero Offset ± 150 mmHg
Probe Impedance 300-3500 ohm input
Frequency Response 0-20 Hz
Input Impedance 820 Kohms @20Hz
>100 Mohms @DC
Transducer Sensitivity 5 µV/V/mmHg (exclusive of transducer accuracy)
Zero Auto push-button
Drift ± 0.1 mmHg/°C
Accuracy ±1 mmHg over full range (exclusive of transducer accuracy)
Scale Ranges -10 to 10, 0 to 20, 0 to 40, 0 to 60, 0 to
100, 0 to 150, 0 to 200, 0 to 300 mmHg
Display Systolic, Diastolic, Mean Pressure
Temperature
Range 20-45° C
68-113° F
Accuracy ± 1° C (20-45° C) (exclusive of transducer accuracy)
Display Resolution 0.1° C or 0.1° F
Probe YSI-400 series compatible
Impedance Respiration
Measurement Method Impedance
Range 0 - 150 breaths per minute
Accuracy Greater of ±1% or ±1 breath/min. over full range
Sensitivity 0.3-10 W (Adult)
0.1-10 W (Neonate)
Oxygen Saturation (%SpO2)
Range 0-99%
Resolution 1%
Accuracy ± 2%, (70-99%)
± 3%, (50-70%)
Method Dual Wavelength LED
Trends
Tabular 24 hours (1 sample/minute)
Memory 24 hours
Display
Type LCD or electroluminescent (EL)
Max. Waveforms 3 simultaneous
Electrical
Power Consumption <80 W Voltage Input 110-220V, 50-60 Hz Classifications
Type of Protection Class I Equipment
Degree of Protection Type CF Defibrillator-Proof Equipment
Protection Against Liquids IPX1, Drip-Proof Equipment
Degree of safety of application in the presence of a flammable
anesthetic mixture with air or with oxygen or nitrous oxide:
Equipment not suitable for use in the presence of a flammable anesthetic
mixture with air or with oxygen or nitrous oxide.
Mode of operation: Continuous operation
Physical Properties
Dimensions 17.3 cm H x 35 cm W x 33 cm D
(6.8" H x 13.8" W x 13" D)
Weight 6.3 kg (13.9 lb.)
Environmental
Operating Temperature 15 to 35°C (59 to 95° F)
Storage Temperature -20 to 60°C (-4 to 140° F)
Humidity 10-90%, noncondensing
Back Panel Outputs and Connectors
Serial Printer or Computer DB-9
Central Station DB-9
Expansion Module DB-9
Data I/O (to/from Data Card) PCMCIA
Analog BNC
Note:
• Performance of the monitor may be adversely affected by exposure to high levels of electrostatic discharge.
|